430
3540
3540
3540

        Downloads


        430 Manual
        3540 Manual
        Catalogue Page - 430
        Catalogue Page - 3540
        Application Notes: pH
        430 Quality Certificate
        3540 Quality Certificate

        Further Information


        Conductivity Cells
        pH Electrodes
        Conductivity Consumables
        pH Buffers
        Dealer Locator
        Request a Quote
        FAQs: Conductivity
        FAQs: pH
        Technical Support
        Warranty
        Service

Click here to Buy Now

Combined pH and Conductivity Meters


430 Enterprise Conductivity/pH Meter

• Simultaneous display of pH, conductivity and temperature
• Automatic or manual calibration
• Combined pH, conductivity and temperature probe
• Power on self check routine
• Auto-selecting ranges for both TDS and conductivity modes

The easy to use 430 is designed to readily withstand the rigours of field work, while reliably measuring pH and conductivity. The unit allows the simultaneous readout of pH, conductivity or TDS and temperature with switched ranges for both conductivity and TDS measurements.

3540 Bench Combined Conductivity/pH Meter

• Simultaneous display of pH and conductivity
• 2 independent channels to prevent interference between probes
• Additional modes for resistivity, salinity, TDS and mV
• Data logger with auto save/print options
• Storage of up to 500 readings (250 for each mode)

The 3540 is ideal for use in all laboratories where pH and conductivity analyses are required. The setup menu gives quick and easy access to the whole range of instrument, pH and conductivity measurement options. Calibration of both channels is automatic, with the option to select 1, 2 or 3 calibration points. The pH channel can also display mV values if required.

Technical Specification

Parameter 430 3540
pH range 0.00 to 14.00 (1 or 2 point calibration) –2.000 to 20.000
pH resolution 0.01 0.001/0.01/0.1
pH accuracy ±0.02 ±0.003
Auto. buffer recognition 4.00, 7.00, 9.22, 10.00 Jenway, NIST, DIN and JIS
mV range ±1999.9mV
mV resolution 0.1mV
mV accuracy ±0.2mV
Conductivity ranges 5 auto–selected from 0 to 199.9mS Auto–selected from 0.01µS to 1.999S
Conductivity resolution 0.01µS to 0.1mS 0.01µS to 1mS
Conductivity accuracy ±0.5% ±2 digits ±0.5% ±2 digits
Cell constant 0.01 to 19.99 0.010 to 19.999
Automatic standard recognition 10µS, 84µS, 1413µS, 12.88mS (with manual override) 10µS, 84µS, 1413µS, 12.88mS (with manual override)
TDS range 5 auto–selected from 0 to 199.9g/l 6 auto–selected from 0 to 1999g/l
TDS resolution 0.01mg/l to 0.1g/l 0.01mg/l to 1g/l
TDS accuracy ±0.5% ±2 digits ±0.5% ±2 digits
EC ratio (TDS) 0.50 to 0.80 0.50 to 0.80
Temperature range –9.9 to 99.9°C –10 to 105°C
Temperature resolution 0.1°C 0.1°C
Temperature accuracy ±0.5°C ±0.5°C
ATC and manual 0 to 99.9°C 0 to 100°C
Temperature coefficient 0.00 to 4.00%/°C 0.00 to 4.00%/°C
Reference temperature 18, 20 or 25°C 18, 20, 25°C
Additional modes Salinity and resistivity
GLP Calibration reminder interval, operator and sample ID and security coded setup
Outputs Analogue, RS232 and IrDA interface
Connector mini–DIN BNC (pH) and 7–pin DIN (conductivity)
Power 2 AA batteries 9V AC ±10% @ 50/60Hz
Size (l x w x d) 175 x 75 x 35mm 210 x 250 x 55mm
Weight 250g 850g


Ordering Information

Part Code Description
430 271 430 conductivitypH meter supplied with combined pHconductivity probe (430 231), buffer sachets, batteries (fitted) and free carry case
354 001 3540 conductivitypH meter supplied with glass combination pH electrode (924 005), glass conductivity probe (K1, 027 013), electrode stand and holder, ATC probe (027 500), BNC shorting plug, pH buffers and UK power supply



Return to Category
Not what you're looking for ? Try our product selector